Excellent condition. Includes 30 day money back guarantee!!!!
ioProfile is a chemistry/cell viability modular instrument that measures up to 15 key cell culture attributes. By combining as many as five separate instruments into one easy-to-use instrument, BioProfile FLEX simplifies workflow and saves time, labor, and operating cost versus multiple instruments.
Applications of BioProfile FLEX include:
- Monitoring the progress of bioreactor runs
- Determining consumption and production of key metabolites
- Identification of growth-limiting nutrients
- Design of feeding strategies
- Calibration of bioreactor probes
- Measurement of cellular respiration
- Balancing media electrolytes
- Control levels of waste product.
- Sample Analysis Time: 2.5 minutes (Gases only – PO2 + PCO2)4.0 minutes
(Chemistry/Gas & Osmolality)6.25 minutes
(All 15 tests)
- Operating Temperature Range: 15-30°C (59-86°F)
- Operating Relative Humidity Range: 20-85%
- Sample Size: 1 mL (for all modules)
- Sampling Options: Individual via syringe/cup
- Automated batch using 20-position sampling tray
- On-line, automated from up to ten bioreactors
- Electrical Requirements: 90-264 VAC, 50 to 60 Hz (Universal Power Supply)
- System Size: Height: 20 in (50 cm) Width: 26 in (66 cm) Depth: 20 in (50 cm)
- System Weight: Analyzer: 94 lbs (42.6 kg) without reagents
- Host Computer Monitor/CPU: 41.5 lbs (18.8 kg)
- Safety Certifications: TUV: IEC 61010-1:2001
- Quality Systems Certifications: TUV: ISO9001:2000 FDA Registered, OPC Compliant, PAT Compatible, 21 CFR Part 11 Compliant
Chemistry/Gas Module AssayMeasurement Range ResolutionMethod
- Glucose 0.2–30.0 g/L*0.01g/L Biosensor
- Lactate 0.2–10.0 g/L*0.01g/L Biosensor
- Glutamine 0.2–12.0 mmol/L*0.01 mmol/L Biosensor
- Glutamate 0.2–12.0 mmol/L*0.01 mmol/L Biosensor
- Ammonium 0.2–25.0 mmol/L0.01 mmol/L Direct ISE
- pH5.000–8.0000.001 Direct ISE
- PCO23.0–300.0 mmHg0.1 mmHg Direct ISE
- PO23.0–800.0 mmHg0.1 mmHg Clarke Electrode
- Sodium 40–220 mmol/L0.1 mmol/L Direct ISE
- Potassium 1.0–100.0 mmol/L0.01 mmol/L Direct ISE
- Calcium 0.10–10.0 mmol/L0.01 mmol/L Direct ISE Calculated Tests:O2 Saturation; CO2 Saturation; Temp. Corrected pH, PCO2, PO2
- *Ranges Reflect User Selectable 1:2 Dilution
- Osmolality ModuleAssayMeasurement Range ResolutionMethod
- Osmolality0–1500 mOsm/kg1 mOsm/kgFreezing PointCell Density/Viability Module AssayMeasurement Range ResolutionMethod
- Cell Diameter8 – 50 μmN/A Digital Imaging
- Detectable Density 50,000 – 100,000,000cells/mL**N/A Digital Imaging
- Viability0 – 100%N/A Digital Imaging Phosphate Module Measurement Range 0.20-10 mmol/L
- Sample Size 500 µL**Ranges Reflect User Selectable 1:2, 1:4, 1:5, 1:10 Dilution
- IgG Module Measurement Range: 0.10–10.00 g/L* *Ranges Reflect User Selectable 1:2 Dilution