- Analytical Instruments
- >
- Advanced Instruments A20 Osmometer
Advanced Instruments A20 Osmometer
Contact for price
BLE2000303
Sample test volume 100μL
Aspirated Sample volume 150μL
Sample capacity 20 sample carousel
Test time Less than 3 minutes
Sample throughput can process 20 samples in less than 1 hour
Unit mOsm/kg H2O
Calibration low: 0 to 2000 mOsm/kg H2O
full: 0 to 4000 mOsm/kg H2O
Resolution1 mOsm/kg H2O
Calibrated 3-point Calibrated for low range
4-point for full range
Communication Onbord printer, 10/100Mbps Ethernet, 4 USB 1.0/1.1/2.0 integrated bar code scanner
Accuracy 0 to 400 mOsm/kg H2O: mean value ≤ 2 mOsm/kg H2O from nominal value 400-4000 mOsm: mean value ≤ 0.5% from normal value
Precision (Within-run) standard deviation ≤ 2 mOsm/kg H2O between 0 and 400 mOsm; standard deviation ≤ 0.5% of value between 400 to 4000 mOsm
Drift Less than 1 mOsm/kg H2O per month
Temperature effects1 Recalibration recommended for ambient temperature change 5°C (9°F) or greater
Storage temperature -40°C to +45°C (-40°F to +113°F)
Simple Viscosity Up to 20 mPa-s(20 cP) higher viscosities may affect pipettor accuracy
Pipettor Accuracy< 1%Electrical voltage100-240 VAC (50/60 Hz)
Power consumption 375W
Dimensions (D x W x H)20.5″ x 23.6″ x 22.8″ (52 cm x 60 cm x 58 cm)
Net weight 68 lb (31 kg)
Shipping weight 133 lb (60 kg)