DIMENSIONS & WEIGHT
Dimensions (W x H x D):790 mm x 1378 mm x 806 mm
(31.1 x 54.25 x 31.73 in)
Weight (Dry):238 kg (525 lbs)
ELECTRICAL
Operating Voltage (older-model power supply; one fan in back cover):103.4 V to 127.2 V (for nominal voltage 110 V to 120 V)
Power Consumption:1200 VA
Operating Voltage (newer-model power supply; two fans in back cover):90 V to 264 V (for nominal voltage 100 V to 240 V)
Mains Frequency:50/60 Hz
ENVIRONMENTAL (UNPACKED INSTRUMENTS)
Maximum Operating Temperature:35°C (95°F)
Minimum Operating Temperature:5°C (41°F)
Temperature Required to meet Staining Performance Requirements:18–26°C (64–79°F)
Maximum Heating Energy Output:1200W/1000W
Operating Humidity (Non-Condensing):30 to 80% RH
Maximum Operating Altitude:0 to 1600 m (5250 ft.) above sea level
Sound Pressure Level Output (at 1 m):< 85 dBA maximum
< 65 dBA normal operation
OPERATING
Slide Capacity:30
Finished trays (10 slides) may be replaced continuously
Number of Reagent Containers:36
Reagent Container Capacity:7 mL and 30 mL
Titration Container Capacity:6 mL
Bulk Reagent Container Capacity:2 L or 5 L
Hazardous Waste Container Capacity:5 L
Standard Waste Container Capacity:2 x 5 L
Chemical Compatibility:All BOND Reagents
70% Alcohol Solution (for cleaning purposes)
Temperature Indication:Defaults (these can be changed by service representatives): Warm: 35°C, Hot: 80°C
Maximum permitted pressure for gas and liquid connections:1.0 bar
MICROSCOPE SLIDES
Slide Width:24.64–26.0 mm (0.97–1.02 in)
Slide Length:74.9–76.0 mm (2.95–2.99 in)
Slide Thickness:0.8–1.3 mm (0.03–0.05 in)
Label Area Width:24.64–26.0 mm (0.97–1.02 in)
Label Area Length:16.9–21.0 mm (0.67–0.83 in)
Material:Glass, ISO 8037/1